1. Intended Use
The Bioperfectus Technologies Influenza A And B Viruses Real Time PCR Kit is an in
vitro diagnostic test, based on real-time PCR technology, for the detection of Influenza A
and B Viruses RNA. Samples can be obtained from throat swab, nasal and pharyngeal
secretions and cell culture supernatant.
2. Kit Components
Components |
Number of Vials |
Volume (μl/Vial) 25T |
Volume (μl/Vial) 50T |
RT-PCR Buffer |
1 |
188 |
375 |
RT-PCR Enzyme Mix |
1 |
125 |
250 |
Reaction Mix |
1 |
188 |
375 |
Positive Control |
1 |
500 |
500 |
Blank Control (RNase-free Water) |
1 |
500 |
500 |
NOTE: Components from different batches can’t be used interchangeably
3. Storage
⚫ All reagents should be stored at -20±5℃ until the expiration date listed on the outer kit box.
⚫ Protect Reaction Mix from light during storage.
⚫ Repeated freezing and thawing (more than five times) of reagents should be avoided.
4. Sample Preparation
⚫ Collect samples in sterile tubes.
⚫ Specimens should be processed immediately or store refrigerated (2-8℃) for no longer than 72 hours before processing. Store any leftover specimens at -20℃ or colder.
⚫ Transportation of clinical specimens must comply with local regulations for the transport of etiologic agents. Transport clinical specimens refrigerated at 2-8℃.
5. Instruction for Use
Learn more, please send us an email: info@bioperfectus.com.